荫道BBWBBB高潮潮喷,涩爱av520com,亚洲免费精品xxxxx,色视频在线免费观看

歡迎來到上海力孚同科技有限公司

服務(wù)熱線:021-64097989

產(chǎn)品展示/ Product display

您的位置:首頁  /  產(chǎn)品展示  /  ECG/EKG Testing  /  The ECG database player  /  MECG2.0ECG database player

ECG database player

ECG database player, MECG 2.0 designed for system validation and software algorithm evaluation. Suitable for R&D and compliance testing. Compliant with IEC 60601-2-25

  • 產(chǎn)品型號:MECG2.0
  • 廠商性質(zhì):代理商
  • 更新時間:2024-09-26
  • 訪  問  量:369
立即咨詢

聯(lián)系電話:021-64097989

詳細介紹

ECG database player, MECG 2.0

The ECG database player designed for system validation and software algorithm evaluation. Suitable for R&D and compliance testing.

Compliant with IEC 60601-2-25, IEC 60601-2-47, ANSI / AAMI EC57, YY0782 and YY0885.

OVERVIEW

Load and replay multi-channel ECG waveforms for diagnostic and ambulatory ECG

Output analog signals simultaneously up to 8 channels and support 12 lead ECG

Embedded CAL, ANE, NST and biological waveforms

Directly support AHA, MIT, CU, NST, ESC and PhysioNet format databases

Support digital file playback function for user to repeatedly play clinical-recorded or programmed waveforms

Extreme output voltage resolution - 0.15μV for MECG 2.0 (2020)

Command replay function enables to program test sequences with commands

Software Development Kit (SDK) allows to develop customized or automated test software with less efforts


SPECIFICATION

ItemsDetails / ReferenceValues

Output channelsThe 8 output channels are provided through a network as specified in IEC 60601-2-51 to provide signals to 10 lead electrodes; in the device under test, this will be displayed as 12 leads.8 outputs

10 lead electrodes

12 leads

Voltage accuracyIEC 60601-2-51 specifies a limit of ±1%, but does not provide a lower limit (all systems must have a lower limit). An inferred specification of 1% ±5µV is derived from the device under test specification in IEC 60601-2-51 of 5% ±25µV.For MECG 2.0, ±1% for values greater or equal to 500μV and ±5μV for values under 500μV.

For MECG 2.0 (2020), ±1% for values greater or equal to 100μV. ±5μV for values under 100μV.

Output voltage resolutionMECG 2.0 uses 12 bit DAC and MECG 2.0 (2020) uses 16 bit DAC.MECG 2.0 for 2.4μV and MECG 2.0 (2020) for 0.15μV

Output voltageThe output voltage on most of the database and ECG is +5mV~-5mV.±5mV

Output noise

level 0-150HzOutput noise should not influence the test. A value a 5µV is suitable for this requirement. Can be verified by monitoring the signal in the device under test using a “diagnostic" filter setting.<5µV

Time accuracyIEC 60601-2-51 does not provide any limits. An inferred limit from the device under test. An inferred limit of ±1% is used. The system’s design accuracy exceeds 0.1% as a 100ppm crystal reference is used.±1%

Sampling rateA maximum sampling rate of 1kHz matches the sampling rates of ECG files.1kHz (8 channels)

Power supplyPowered from the USB supply (5V 0.2A)N/A

EnvironmentIntended for normal laboratory environment. The selection of critical components such as reference voltages, DAC, precision resistors are known to be stable in the range shown.15 - 30°C

10 - 95% RH


外形尺寸: 16.5cm(長)*9.8cm(寬)*4.2cm(高),接線盒尺寸:13.5cm(長)*17.5cm(寬)*4.5cm(高)




留言詢價

留言框

  • 產(chǎn)品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細地址:

  • 補充說明:

  • 驗證碼:

    請輸入計算結(jié)果(填寫阿拉伯數(shù)字),如:三加四=7

C

CODE
關(guān)注我們
  • 聯(lián)系電話:13671818969

  • 聯(lián)系郵箱:lisa815@yeah.net

  • 公司地址:中國(上海)自由貿(mào)易試驗區(qū)臨港新片區(qū)老蘆公路536號

Copyright © 2024 上海力孚同科技有限公司版權(quán)所有   備案號:   技術(shù)支持:化工儀器網(wǎng)

sitmap.xml   管理登陸